Main performance characteristics and parameters:

  1. ★Adopt ≥15-inch color LCD touch screen as the user interface, all in Chinese.
  2. • Environmental protection: The tissue treatment system is in a closed state, with independent liquid and gas passages, and the gas in the air passage can be liquefied and discharged into the waste liquid bottle through the uniquely designed condenser tube, without the leakage of polluting gas, which meets the environmental protection requirements.
  3. • The sample is not moved, and the reagent is moved in the sample processing mode, which can avoid the risk of damage to the sample caused by other mechanical failures such as power failure of the dehydrator.
  4. • Timing function: You can set the completion time of dehydration on any day of the week.
  5. • With the function of power failure protection, the instrument can automatically work from the breakpoint in sequence after the UPS power supply is automatically activated after a power failure or when the power is turned on again.
  6. • There are ≥10 sets of tissue processing programs for users to set and choose.
  7. ★Reagent management system: Set the use cycle of reagents and paraffin according to the number of cassettes processed, the days of reagent use or the number of dehydration times. The system automatically counts, and automatically prompts to replace the reagents and paraffin when the threshold is reached.
  8. ★It has the function of rapid dehydration, which can quickly dehydrate small specimens with conventional reagents, and complete dehydration within 3 hours.
  9. ★Gas pressure conversion device: The gas pressure conversion that used to rely on multiple solenoid valves is now completed by a mechanical structure device, which avoids the high failure rate caused by solenoid valves.
  10. ★Automatic rotation function of paraffin or reagent: According to the effective use threshold of paraffin or reagent, it will automatically remind to replace when the threshold is reached. The system will automatically rotate from low to high according to the cleanliness, waste wax or waste liquid will be automatically discharged to the drainage box, and new wax or new reagent can be added to the emptying position.
  11. ★Multiple safety protection functions: 2 optical sensors sense the lowest and highest liquid level, prompt and stop running protection functions. Studio over-temperature detection and power failure protection. The studio cover is not locked with automatic alarm function.
  12. • With automatic remote telephone alarm function, timely check equipment failure (optional).
  13. ★Stirring function: It has various stirring methods of normal pressure, pressurization, vacuum and magnetic stirring. You can set the stirring time, interval and whether to turn on the stirring function.
  14. • Processes up to 300-350 standard cassettes in an ordered sequence at a time.
  15. • Number of Reagent Bottles: 13
  16. • Number of wax baths: 3, number of working cylinders: 1
  17. • Studio volume: 9L, reagent bottle volume: 6L
  18. • Working room temperature: when the medium is solvent, the temperature is ≤45℃, and when the medium is paraffin, the temperature is 58℃-70℃, which can be adjusted.
  19. • Working pressure: positive pressure <0.03Mpa negative pressure 0-70Kpa, adjustable.
  20. • Wax bath temperature: 58℃-70℃, adjustable
  21. • Paraffin melting time≤3 hours
  22. • Bath setting time: 0~300min, adjustable
  23. • Inlet time: ≤5min; discharge time ≤5min
  24. • Stirring time: 2min
  25. • Stirring interval: 8min (preset)