Pulmonary Function Tester A9
Pulmonary Function Tester A9
Intended Use
Pulmonary function tester A9 is a professional system for detecting lung diseases. lt consists of a portable pulmonary function tester host and application software. lt can conduct comprehensive pulmonary ventilation function tests, including forced vital capacity, vital capacity, maximum Voluntary Ventilation, and diastolic medication before and after comparison tests.
The test is conducted under the guidance of a doctor in a clinic, doctor's office, or hospital. it is suitable for people over 4 years old who can cooperate with the test. it can be used for the detection and diagnosis of bronchial asthma, chronic
obstructive pulmonary disease and other diseases.
Highlights
- The differential pressure flow sensor has high accuracy, good repeatability, stable performance, long service life, simple maintenance, and low cost;
- Supports USB and Bluetooth data transmission methods, and can freely choose PC or APP test software according to the usage scenario;
- Automatically give ATS2019 quality control ratings, intelligently interpret lung function reports, and provide analysis suggestions;
- Breathing tubes can be completely immersed and sterilized, reducing cross-infection, and is compatible with common bacterial filters.
Complete, Professional, and Flexible Testing System
- Multi-mode and multi-parameters: Forced Vital Capacity (FVC), Slow Vital Capacity (SVC), Maximum Voluntary Ventilation (MVV), bronchodilation test, with more than 30 measurement parameters;
- Supports capacity/3 flow calibration and verification to ensure the accuracy and stability of measurement data;
- Supports multiple prediction models, including adult prediction (ECCS93), children prediction (Zapletel), etc.;
- Supports standard follow-up questionnaires for common respiratory diseases: including CAT, mMRC, COPD-SQ, etc.
Comprehensive Quality Control of Pulmonary Function Tests
- Standardized quality control of operations: real-time monitoring of forced exhalation duration, guidance on stopping exhalation when forced end-expiration flow is lower than 0.025L/s, guidance on zero point calibration of pressure difference, and a dynamic reminder of tidal baseline;
- Automatic quality control rating: A/B/C/D/E/U/F;
- Result interpretation: Intelligent interpretation of pulmonary function reports, automatically evaluates test results and gives disease analysis suggestions, standardized clinical scales assist diagnosis and treatment decisions.
Technical Parameters
Flow |
Range: 0-16L/s; Accuracy: <5% |
Volume |
Range: 0-10L; Accuracy :<2.5% |
Resistance |
<0.15kPa/(L/s) |
Test parameters |
FVC、 FEVO.5、 FEV1、 FEV3、 FEV6、 V backextrapol.ex、 FIVC、 |
Optional accessories
3L Calibration Cylinder |
|
Disposable Filters |
|